DMTA-10084-01EN, Rev. 2, November 2018
Chapter 11
196
Figure 11-35 Transducer Setup page
10. Press ESC twice to dismiss the Transducer Setup page and display the Do-Zero
dialog box (see Figure 11-36 on page 196).
Figure 11-36 Do-Zero dialog box
11. Press the Enter key to perform a Do-Zero and begin the calibration.
11.7.3.1 Improving Measurement Accuracy
The combination of default settings from the automatic probe recognition and the
do-zero function enable you to immediately begin measuring with a reasonable
amount of accuracy.
You can improve upon the do-zero measurement accuracy by performing a straight
beam thickness calibration while the corrosion module is active. For best results,
Olympus recommends that you perform a separate calibration in both modes