DMTA-10084-01EN, Rev. 2, November 2018
Gates
73
Figure 7-2 Gate 1 and gate 2 (with echo-to-echo turned on)
Both gates can be used to do the following:
• Measure thickness with straight-beam transducers
• Measure sound path and depth with angle beam transducers
• Measure signal amplitude
• Measure time-of-flight in microseconds
• Trigger threshold and minimum-depth alarms
The gates can also be used together to take echo-to-echo thickness measurements.
7.3 Quickly Adjusting Basic Gate Parameters
You can perform basic gate adjustments using the Gate1 and Gate2 sidebar submenus
(see Figure 7-3 on page 74).
Gate 1 (red)
Gate 2 (blue)