DMTA-10084-01EN, Rev. 2, November 2018
Software Features and Options
169
4. Turn the adjustment knob to highlight the Add button, and then press the Enter
key to add the scanning gain to the base gain and to apply the transfer correction.
11.3.6 JIS DAC
The Japanese Industrial Standard (JIS) DAC mode meets the requirements of JIS
Z3060. The JIS DAC curve setup is nearly identical to the standard DAC/TCG setup.
However, any of the six curves can be used to trigger the alarm when in the JIS DAC
mode. Additionally, you can set the alarm to positive or negative.
To select the curve to be used as the alarm reference level
1. Activate the DAC/TCG feature (see “Enabling the DAC/TCG Feature” on
page 154).
2. In the DAC/TCG dialog box, turn the adjustment knob to set the DAC/TCG
Mode to JIS (see Figure 11-15 on page 169).
Figure 11-15 JIS mode
3. Press the Shift, Tab key sequence to return to the live A-scan display.