Figure 11-19 DGS/AVG dialog box
3. Press the Enter key to highlight the DGS/AVG parameter box.
4. Turn the adjustment knob to set the DGS/AVG Mode to On.
The DGS/AVG dialog box enables you to define the exact probe being used for
the inspection, and to set up the DGS/AVG curve to be drawn. The adjustments
that can be made are as follows:
— Probe Type
Selects the type of probe to be used (Straight Beam, Angle Beam, or Dual
Probe).
— Name
Selects the actual probe to be used.
—Refl. Type
Defines the type of reflector to be used to acquire a reference indication to
build the DGS/AVG curve. For straight beam and dual probes, the available
reflectors are as follows:
○ Back wall
○ Side-drilled hole (SDH)
○ Flat-bottom hole (FBH)