DMTA-10084-01EN, Rev. 2, November 2018
Software Features and Options
161
Figure 11-7 Partial DAC curve with each echo set to 80 % FSH
11.3.3.1 Other Adjustments for Building a DAC Curve
You can increase the instrument range or increase the display delay to increase the
number of echoes that you can see.
To adjust the Range, Delay, or Zero while detecting the echo
1. On the DAC/TCG submenu, press the ESC key to go to the sidebar menu.
2. Turn the adjustment knob to select the UT button, and then press the Enter key to
display the UT submenu.
3. Adjust the Range, Delay, or Zero parameters while detecting the echo.
To adjust the Angle, Thickness, and Curved Surface Correction values
1. On the sidebar menu, press the Shift key, and then press the ESC key to go to the
setup menu.
2. Turn the adjustment knob to select the Angle icon, and then press the Enter key to
open the Angle dialog box.