DMTA-10084-01EN, Rev. 2, November 2018
Data Logger
113
To take a snap shot of the waveform
1. With the waveform data page displayed, turn the adjustment knob to highlight
the Snap Shot button (see Figure 9-9 on page 113).
2. Press the Enter key to save the snap shot to the cloud or a USB flash drive as a
bitmap (.bmp) image file.
Figure 9-9 Saved file waveform and data
9.4.2.4 Recalling a File ID
To bring saved data to the live A-scan display, you must recall that particular file ID.
For incremental (INC) files, you must select the specific ID to be recalled. For
calibration (CAL) files, recalling the file automatically recalls the parameters of the
single ID saved in that file.