DMTA-10084-01EN, Rev. 2, November 2018
Calibration
97
Figure 8-5 Calibration Ready to collect thin echo
10. Couple the transducer to the 0.400 in. step of the calibration test block.
11. Repeat the Auto 80% procedure for this step of the calibration test block.
12. Press the Tab key to highlight the Collect button (see Figure 8-6 on page 98).