Olympus EPOCH 6LT - Figure 8-4 Calibration - Ready to Auto 80% on Thin Echo

Olympus EPOCH 6LT
232 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DMTA-10084-01EN, Rev. 2, November 2018
Chapter 8
96
Figure 8-4 Calibration — Ready to Auto 80% on thin echo
9. Press the Tab key to highlight the Collect button, and then press the Enter key to
capture the thin echo (see Figure 8-6 on page 98).