DMTA-10084-01EN, Rev. 2, November 2018
Data Logger
115
To recall a file ID from the File setup page
1. In the File setup page, turn the adjustment knob, and then press the Enter key to
select the file that you want to recall.
2. Press the Tab key to select the Contents button.
3. Press the Enter key to view the file contents.
4. Press the Tab key to select the Recall button.
5. Press the Enter key to bring the saved data to the live A-scan display.
The ID being viewed is displayed in the file identifier bar. The notification
“<filename> recalled, press any key to continue” is displayed in the message bar
(seeFigure 9-10 on page 114).
6. Press any key on the instrument to continue.
9.4.2.5 Viewing File Details
You can view the file setup and creation information.
To view the file details
1. In the File setup page, turn the adjustment knob, and then press the Enter key to
select the file that you want to use.
2. Press the Tab key to select the Details button.
3. Press the Enter key to view the Details page (see Figure 9-11 on page 116).