DMTA-10084-01EN, Rev. 2, November 2018
Data Logger
129
Copy language BIN file to USB mass storage
Exports current language file from the internal memory to the USB flash
drive.
Copy XML Schema File from USB mass storage
Imports an XML Schema file from the USB flash drive to the internal memory
to support file exports in the XML format.
Copy XML Schema File to USB mass storage
Exports the current XML Schema file from the internal memory to the USB
flash drive.
Copy DGS Custom Probe File from USB mass storage
Imports new custom DGS/AVG probe library from the USB flash drive to the
internal memory.
Copy DGS Custom Probe File to USB mass storage
Exports current DGS/AVG custom probe library from the internal memory to
the USB flash drive.
Restore database files from USB mass storage to internal
Restores all data logger files and live instrument settings from the USB flash
drive to the internal memory from a previously backed up file.
Backup database files from internal to USB mass storage
Backs up all data logger files and live instrument settings from the internal
memory to a USB flash drive.
Restore all files from USB mass storage to internal
Restores all data logger files and live instrument settings from the USB flash
drive to the internal memory from a previously backed up file.
Backup all files from internal to USB mass storage
Backs up all data logger files and live instrument settings to a single file on
the USB flash drive for data archiving or instrument cloning.
3. Press the Tab key to highlight Start, and then press the Enter key to begin the
transfer.
A dialog box appears, asking you to confirm the transfer.
4. Press the Enter key to confirm the transfer.