Figure 2-25: Fundamental wave and the frequency and level reference point
5. Make the step size for the center frequency equal to the signal frequency
a) Press the FREQ key.
b) Press the "CF-Stepsize" softkey and press the "= Marker" softkey in the sub-
menu.
The step size for the center frequency is now equal to the marker frequency.
6. Set the center frequency to the second harmonic of the signal.
a) Press the FREQ key.
b) Press the UPARROW key once.
The center frequency is set to the second harmonic.
7. Place the delta marker on the second harmonic.
a) Press the MKR -> key.
b) Press the "Peak" softkey.
The delta marker moves to the maximum of the second harmonic. The displayed
level result is relative to the reference point level (= fundamental wave level).