Instrument Functions
R&S
®
FSVA/FSV
346Operating Manual 1176.7510.02 ─ 09
Save As Standard.........................................................................................355
Meas Start/Stop............................................................................................ 355
Restore Standard Files................................................................................. 356
Spectrum Emission Mask
Opens a submenu to configure the Spectrum Emission Mask measurement.
The Spectrum Emission Mask (SEM) measurement defines a measurement that moni-
tors compliance with a spectral mask. The SEM measurement of the base unit allows a
flexible definition of all parameters in the SEM measurement.
For general information on performing SEM measurements, see Chapter 3.3.5.5,
"Measuring with Spectrum Emission Masks", on page 344.
Remote command:
SENS:SWE:MODE ESP, see [SENSe:]SWEep:MODE on page 857
Sweep List ← Spectrum Emission Mask
Opens a submenu to edit the sweep list and displays the "Sweep List" dialog box.