Figure 3-33: Measurement with Reference Value: +10dB and Reference Value Position = 50%
After the reference line has been shifted by entering +10 dB as the "Reference Value"
on page 477, deviations from the nominal value can be displayed with high resolution
(e.g. 2 dB/Div.). The absolute measured values are still displayed; in the above exam-
ple, 2 dB below nominal value (reference line) = 8 dB attenuation.
Instrument Functions – Tracking Generator (Options R&S FSV-B9/ R&S FSV-B10)