Instrument Functions
R&S
®
FSVA/FSV
536Operating Manual 1176.7510.02 ─ 09
Test Setup
Figure 3-45: External mixer test setup
1. 1. Connect the "LO OUT / IF IN" output of the R&S FSVA/FSV to the LO/IF port of
the external mixer.
2. Connect the multiplier to the external mixer.
3. Apply a sine wave signal with f = 14.5 GHz to the input of the multiplier.
Activating the External Mixer and Selecting the Mode
► Activate the external mixing for all applications by selecting "FREQ > External
Mixer > External Mixer ON" in Spectrum mode.
Default Settings
Prior to the measurement select the required band (in this case the V band).
1. Select "FREQ > External Mixer > External Mixer Config".
2. From the "Band" selection list, select the "V" band.
Instrument Functions – External Mixer (Option R&S
FSV-B21)