Remote Control
R&S
®
FSVA/FSV
757Operating Manual 1176.7510.02 ─ 09
In single sweep mode, you can synchronize to the end of the measurement with *OPC,
*OPC? or *WAI. In continuous sweep mode, synchronization to the end of the mea-
surement is not possible. Thus, it is not recommended that you use continuous sweep
mode in remote control, as results like trace data or markers are only valid after a sin-
gle sweep end synchronization.
Suffix:
<n>
.
irrelevant
Example:
INIT:CONT OFF
Switches to single sweep mode.
DISP:WIND:TRAC:MODE AVER
Switches on trace averaging.
SWE:COUN 20
Setting the sweep counter to 20 sweeps.
INIT;*WAI
Starts the measurement and waits for the end of the 20 sweeps.
INIT:CONM;*WAI
Continues the measurement (next 20 sequences) and waits for
the end.
Manual operation: See "Continue Single Sweep" on page 247
INITiate<n>:CONTinuous <State>
This command determines whether the trigger system is continuously initiated (contin-
uous) or performs single measurements (single).
The sweep is started immediately.
In the "Spectrum" mode, this setting refers to the sweep sequence (switching between
continuous/single sweep).
In single sweep mode, you can synchronize to the end of the measurement with *OPC,
*OPC? or *WAI. In continuous sweep mode, synchronization to the end of the mea-
surement is not possible. Thus, it is not recommended that you use continuous sweep
mode in remote control, as results like trace data or markers are only valid after a sin-
gle sweep end synchronization.
Suffix:
<n>
.
irrelevant
Parameters:
<State> ON | OFF
*RST: ON
Example:
INIT:CONT OFF
Switches the sequence to single sweep.
INIT:CONT ON
Switches the sequence to continuous sweep.
Mode: all
Manual operation: See "Continuous Sweep" on page 246
See "Single Sweep" on page 247
Remote Control – Commands