Remote Control
R&S
®
FSVA/FSV
882Operating Manual 1176.7510.02 ─ 09
Parameters:
<GPIBAddress>,
<Command>
<0…30>: GPIB address of the external generator connected via
the GPIB interface.
<command>: String with the command. Only basic I/O is possi-
ble.
Example:
"SYST:COMM:GPIB:RDEF:COMM 18,'*RST'"
Resets generator.
"SYST:COMM:GPIB:RDEF:COMM 18,'SOURCE:FREQ:CW
2E9'"
Sets generator frequency to 2 GHz.
"SYST:COMM:GPIB:RDEF:COMM 18,'SOUR:POW 0'"
Sets generator power to 0dBm.
"SYST:COMM:GPIB:RDEF:COMM? 18,'SENS:POW?'"
Queries the generator power.
Mode: A
SYSTem:COMMunicate:GPIB:RDEVice:GENerator<generator>:ADDRess
<Number>
Changes the IEC/IEEE-bus address of the external tracking generator.
This command is only available if the R&S FSVA/FSV option B10 (External Tracking
Generator) is installed.
Suffix:
<generator>
.
1 | 2
tracking generator
For Noise Figure measurements (K30): 1 only
Parameters:
<Number> Range: 0 to 30
*RST: 28
Example:
SYST:COMM:GPIB:RDEV:GEN1:ADDR 15
Mode: A, NF
SYSTem:COMMunicate:RDEVice:GENerator<generator>:INTerface <Type>
Defines the interface used for the connection to the external tracking generator.
This command is only available if the R&S FSVA/FSV option B10 (External Tracking
Generator) is installed.
Suffix:
<generator>
.
1 | 2
tracking generator
For Noise Figure measurements (K30): 1 only
Remote Control – Commands