Remote Control
R&S
®
FSVA/FSV
952Operating Manual 1176.7510.02 ─ 09
Manual operation: See "Edit" on page 175
[SENSe:]CORRection:TRANsducer:DATA <Frequency>,<Level>
This command defines the shape of the transducer factor.
A transducer factor may contain up to 625 reference values. Each reference value is a
combination of a frequency and a level value.
Before you can use this command, you have to select a transducer factor with
[SENSe:]CORRection:TRANsducer:SELect on page 953.
Parameters:
<Frequency> Numeric value in Hz.
The frequency values have to be in ascending order.
<Level> The unit depends on [SENSe:]CORRection:TRANsducer:
UNIT on page 953. By default, it is dB.
Example:
SENS:CORR:TRAN:SEL 'TRAN_1'
Selects the transducer factor
CORR:TRAN:UNIT 'DB'
Selects the unit for the transducer.
CORR:TRAN:DATA 1MHZ,-30,2MHZ,-40
Defines the frequency/level pairs 1 MHz, -30 dB and 2 MHz, -40
dB.
Manual operation: See "Edit Value" on page 175
[SENSe:]CORRection:TRANsducer:DELete
This command deletes a transducer factor.
Before you can use the command, you have to select a transducer factor with
[SENSe:]CORRection:TRANsducer:SELect.
Example:
CORR:TRAN:SEL 'Transducer 1'
CORR:TRAN:DEL
Deletes the selected transducer.
Usage: Event
Manual operation: See "Delete" on page 176
[SENSe:]CORRection:TRANsducer:SCALing <Scale>
This command selects the scale of the transducer factor.
Before you can use the command, you have to select a transducer factor with
[SENSe:]CORRection:TRANsducer:SELect on page 953R&S FSVA/FSV.
Parameters:
<Scale> LINear | LOGarithmic
*RST: LINear
Remote Control – Commands