Proceed as follows:
●
Open the R&S website http://www.rohde-schwarz.com in section "Power Meters &
Voltmeters > R&S NRP-Z Sensors".
●
Select the respective sensor, e.g. R&S NRP-Z81.
●
Select "Downloads > Firmware" and the offered firmware suitable for your sensor.
●
Save the firmware in the /var/user/ directory.
●
Connect the sensor to the R&S SMB and select "Setup > System > NRP-Z Info
Update" to open the dialog.
●
Select the sensor in the left sensor selection field.
●
Select the update file with "Select NRP File" and activate "Run Update".
The update starts and a bar indicates the progress.
Note: If the update is interrupted for example by accidental removal of the sensor during
the process, the button "Rescue" appears. Thus, you can restart the update process.
Prerequisite is that no other sensor is connected to the instrument.
The procedure is as follows:
●
Do not reconnect the sensor but keep it ready to be connected
●
Select "Rescue" in the left sensor selection field
●
Activate "Run Update"
●
Confirm query in message box
●
Connect sensor within 4 seconds
The update starts, a bar informs about the progress.
Remote command:
n.a.
5.2.3.7 SMZ Info Update
The "SMZ Info/Update" dialog covers information on a connected frequency multiplier,
like type, serial number revision state, frequency range and firmware version. You can
directly perform an update of the multiplier firmware.
To access the "SMZ Info/Update" dialog, perform one of the following
●
On the front panel, press the SETUP key and select "SMZ > SMZ Info Update".
●
In the block diagram, select "RF Block > Configure > SMZ Info Update"