Remote Control Commands
R&S
®
SMB100A
299Operating Manual 1407.0806.32 ─ 13
Example:
CORR:CSET '/var/user//Lists/ucor/ucor1'
selects the table ucor1.
CORR ON
activates level correction. Correction is performed using the table
ucor1.
Manual operation: See "User Cor. Data - User Correction" on page 135
[:SOURce<hw>]:CORRection[:STATe] <State>
The command activates/deactivates level correction. Level correction is performed using
the table which has been selected with the command CORRection:CSET:SELect.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
SOUR:CORR:CSET '/var/user//lists/ucor/ucor1'
selects the table ucor1.
SOUR:CORR ON
activates user correction.
Manual operation: See "State - User Correction" on page 135
[:SOURce<hw>]:CORRection:VALue?
The command requests the current value for user correction.
Return values:
<Value> float
Range: -100 to 100
Increment: 0.01
*RST: 0
Example:
CORR:VAL?
queries the value currently used for level correction.
Response: -3
the correction value is - 3 dB.
Usage: Query only
Manual operation: See "User Correction Value - User Correction" on page 135
7.12.3 SOURce:FM Subsystem
The FM subsystem contains the commands for checking the frequency modulation.
Characteristics which are valid for all modulations and the LF Output are configured in
the SOURce:LFOutput subsystem (e.g. frequency). The external signal is input at the
MOD EXT connector.