Remote Control Commands
R&S
®
SMB100A
356Operating Manual 1407.0806.32 ─ 13
Return values:
<Points> integer
Range: 0 to INT_MAX
*RST: 0
Example:
MMEM:CDIR '/var/user/lists/pulsetrain'
selects the directory for the pulse train files.
PULM:TRA:SEL ‘P_INCR'
selects P_INCR for editing. P_INCR is created if it does not yet
exist.
PULM:TRA:OFFT:POIN?
queries the number of frequency values in P_INCR
Response: 7
P_INCR has 7 offtime entries.
Usage: Query only
Options: R&S SMB-K27 (Pulse Train)
[:SOURce<hw>]:PULM:TRAin:ONTime <OnTime>
Fills the Ontime part of the selected file with data.
Parameters:
<OnTime> Onime#1{, Ontime#2, ...} | binary block data
The data can be given either as a list of numbers (list can be of
any length and list entries must be separated by commas) or as
binary block data.
When block data is transferred, 8 (4) bytes are always interpreted
as a floating-point number with double accuracy (see the com-
mand FORMat:DATA).
The maximum length is 2047 values.
Range: 0 ns to 5 ms
Example:
MMEM:CDIR '/var/user/lists/pulsetrain'
selects the directory for the pulse train files.
PULM:TRA:SEL ‘P_INCR'
selects P_INCR for editing. P_INCR is created if it does not yet
exist.
PULM:TRA:ONT 10ns,15ns,40ns,...
specifies the ontime values in P_INCR. If the list already contains
data, it is overwritten.
Options: R&S SMB-K27 (Pulse Train)
Manual operation: See "Edit Pulse Train Data" on page 198
[:SOURce<hw>]:PULM:TRAin:ONTime:POINts?
Queries the length (in points) of the ontime component of the selected list.
SOURce Subsystem