Remote Control Commands
R&S
®
SMB100A
364Operating Manual 1407.0806.32 ─ 13
Parameters:
<Mode> IMPort | EXPort
*RST: IMPort
Example:
PULM:TRA:DEXC:MODE IMP
selects that ASCII files with ontime/offtime/repetition values are
imported and transferred into pulse train lists.
MMEM:CDIR 'var/lists/pulsetrain/import'
selects the directory for the ASCII files with ontime/offtime/repeti-
tion values.
PULM:TRA:DEXC:AFIL:SEL 'train.csv'
selects that ASCII file train.csv is imported.
PULM:TRA:DEXC:SEL 'train_imp'
selects that the ASCII file train.csv is imported into pulse train
list train_imp.
Options:
R&S SMB-K27 (Pulse Train)
Manual operation: See "Mode – Import/Export Pulse Train Files" on page 200
[:SOURce<hw>]:PULM:TRAin:DEXChange:SELect <Filename>
The command selects the pulse train list to be imported or exported.
The pulse train files are stored with the fixed file extensions *.pulstrn in a directory of
the user's choice. The directory applicable to the commands is defined with the command
MMEMory:CDIR.
Parameters:
<Filename> string
Example:
PULM:TRA:DEXC:MODE IMP
selects that ASCII files with ontime/offtime/repetition values are
imported and transferred into pulse train lists.
MMEM:CDIR 'var/lists/pulsetrain/import'
selects the directory for the ASCII files with ontime/offtime/repeti-
tion values.
PULM:TRA:DEXC:AFIL:SEL 'train.csv'
selects that ASCII file train.csv is imported.
PULM:TRA:DEXC:SEL 'train_imp'
selects that the ASCII file train.csv is imported into pulse train
list train_imp.
Options:
R&S SMB-K27 (Pulse Train)
Manual operation: See "Select Destination / Source - Import/Export Pulse Train
Files" on page 201
[:SOURce<hw>]:PULM:WIDTh <Width>
Sets the width of the generated pulse. The width determines the pulse length. The pulse
width must be at least 20ns less than the set pulse period.
SOURce Subsystem