Instrument Function
R&S
®
SMB100A
119Operating Manual 1407.0806.32 ─ 13
Variation Step - RF Signal
Sets the user-defined step width. This step width is used when entering the RF frequency
using the rotary knob. Frequency variation with this step width must also be activated
with "Variation Active".
Remote command:
[:SOURce<hw>]:FREQuency:STEP[:INCRement] on page 309
5.3.3 Phase
The phase of the RF output signal can be changed in the "Phase Settings" section of the
"RF Frequency/Phase" dialog.
5.3.3.1 Phase Settings
To open the "RF Frequency/Phase" menu, select "RF" > "Configure" >"Frequency /
Phase" or use the MENU key under "RF".
Delta Phase - RF Signal
Sets the phase of the RF signal. The current phase of the signal is used as the reference.
This function allows, for example, the phase of the output signal to be synchronized with
the phase of a signal from a second signal generator.
Remote command:
[:SOURce<hw>]:PHASe on page 334
Reset Delta Phase Display - RF Signal
Resets delta phase value. The set phase is adopted as the new current phase, i.e. the
delta phase value is reset to 0.
Remote command:
[:SOURce<hw>]:PHASe:REFerence on page 335
5.3.4 Reference Oscillator
In the internal reference mode the internal reference signal is available at the REF
OUT connector. The frequency of the internal reference signal is permanently set to 10
MHz.
The settings of the reference oscillator are not influenced by an instrument preset
("PRESET" key).