Example:
SENS1:FILT:TYPE NSR
selects fixed noise filter mode for the power sensor connected to
the SENSOR connector.
SENS1:FILT:NSR .2
sets a noise content of 0.2.
SENS1:FILT:NSR:MTIM 5
limits the settling time to 5 seconds
Manual operation: See "Filter Length - Power Sensors" on page 145
:SENSe<ch>[:POWer]:FILTer:SONCe
The command activates the search for the optimum filter length for the current measure-
ment conditions. The found filter length can be retrieved with com-
mand :SENSe:POWer:FILTer:LENGth:USER?. This command is only available for
user filter mode (:SENSe:POWer:FILTer:TYPE USER).
Example:
SENS:FILT:TYPE USER
selects user filter mode.
SENS:FILT:SONC
activates the search for the optimum filter length.
SENS:FILT:LENG?
returns the found optimum filter length.
Response: 128
Usage: Event
Manual operation: See "Auto Once" on page 145
:SENSe<ch>[:POWer]:FILTer:TYPE <Type>
The command selects the filter mode. The filter length is the multiplier for the time window