Remote Control Commands
R&S
®
SMB100A
405Operating Manual 1407.0806.32 ─ 13
:TEST<hw>:ALL:RESult?
Queries the result of the performed selftest (command :TEST<hw>:ALL:STARt
on page 404).
Return values:
<Result> 0 | 1 | RUNning | STOPped
*RST: STOPped
Example:
TEST:ALL
Starts the selftest
:TEST:ALL:RES?
Response: "0" on success, "1" on fail
Usage: Query only
Manual operation: See "Selftest" on page 89
:
TEST<hw>:DIRect <HW_assembly>,<subadress>,<hex data string>
:TEST<hw>:DIRect? <HW_assembly>,<subadress>
The respective hardware assembly responds directly to the command; any safety mech-
anisms are bypassed. This function is only available via remote control.
Example:
TEST:DIR 'SSYN',0,#H12345678
TEST:DIR? 'SSYN',0
Response: #H12345678
7.16 TRIGger Subsystem
The TRIGger system contains the commands for selecting the trigger source for the RF
and LF sweep. The trigger input connectors are configured in the SOURce:INPut sub-
system.
The trigger system of the R&S SMB is a simplified implementation of the SCPI trigger
system. The TRIGger system differs from the SCPI system as follows:
No INITiate command; the instrument behaves as if INITiate:CONTinuous
ON were set.
Under TRIGger several sweep subsystems exist.
Other commands associated with the trigger system of the R&S SMB can be found in the
modulation and RF signal subsystems.
TRIGger<hw>
Suffix TRIGger<1|2> is not permitted
TRIGger0 activates the LF output.
TRIGger Subsystem