:SYSTem:COMMunicate:NETWork[:IPADdress]:SUBNet:MASK <Mask>
Sets the Subnet mask. The entry is only activated after sending command
SYST:COMM:NETW:IPAD:SET.
Parameters:
<Mask> string
Example:
SYST:COMM:NETW:SUBN:MASK 255.255.0.0
sets the Subnet mask.
SYST:COMM:NETW:IPAD:SET
activates all changes of the IP address settings.
Manual operation: See "Subnet Mask" on page 92
:SYSTem:COMMunicate:NETWork:RESource?
Queries the VISA resource string. This string is used for remote control of the instrument
with VXI-11 protocol.
Return values:
<Resource> string
Example:
SYST:COMM:NETW:RES
queries the VISA resource string.
Response: TCPIP::192.1.2.3::INSTR
Usage: Query only
Manual operation: See "Visa Resource Strings" on page 96
:SYSTem:COMMunicate:USB:RESource?
Queries the visa resource string for remote control via the USB interface.
Return values:
<Resource> string
Example:
SYST:COMM:USB:RES?
queries the VISA resource string for remote control via the USB
interface.
Response: USB::72::000000::INSTR
Usage: Query only
Manual operation: See "Visa Resource Strings" on page 96
:SYSTem:COMMunicate:SERial:RESource?
Queries the visa resource string for the serial remote control interface. This string is used
for remote control of the instrument.
Return values:
<Resource> string
SYSTem Subsystem