Instrument Function
R&S
®
SMB100A
155Operating Manual 1407.0806.32 ─ 13
"Negative"
The falling edge of the trigger signal is active.
Remote command:
[:SOURce]:INPut:TRIGger:SLOPe on page 310
5.3.7.3 RF Level Sweep
To open the "Level Sweep" menu, select "RF > Configure > Level Sweep" or use the
MENU key under "RF".
RF Level Sweep Settings
In the top section, the RF level sweep is activated and the sweep mode is selected. The
buttons are used to reset the level sweep (all sweep modes) or to execute the level sweep
("Single" mode).
The sweep range, sweep spacing and dwell time are set in the bottom section.
State - Level Sweep
Activates Level Sweep mode.
Note:
Activating a sweep mode automatically deactivates other sweeps and the list mode.
Remote command:
[:SOURce<hw>]:POWer:MODE on page 345
Mode - Level Sweep
Selects the level sweep instrument operating mode and the sweep mode.
If you change the sweep mode during the execution, the signal generator stops the sweep
and starts with the next trigger event at the initial value.
The "Reset Sweep" button sets the sweep to the start value.
RF Block