Remote Control Commands
R&S
®
SMB100A
411Operating Manual 1407.0806.32 ─ 13
Manual operation: See "Mode - LF Frequency Sweep" on page 190
:TRIGger<hw>[:SWEep][:IMMediate]
The command starts all sweeps which are activated for the respective path. The com-
mand starts all sweeps which are activated.
The sweep to be executed depends on the respective MODE setting
(:SOUR:SWEep:POW|FREQ:MODE and :SOUR:LFO:SWEep[:FREQ]:MODE).
The command corresponds to the manual-control command "Execute Trigger".
Example:
TRIG
starts all active sweeps.
Usage: Event
Manual operation: See "Execute Single Sweep - Frequency Sweep" on page 151
:
TRIGger<hw>[:IMMediate]
The command immediately starts the activated sweep.
The command performs a single sweep and therefore applies to sweep mode AUTO with
sweep source SINGle. Use the commands TRIG:FSW|LFFS|PSW|[:SWE]:SOUR
SING, and SOUR:SWE:FREQ|POW:MODE, or SOUR:LFO:SWE:[FREQ:]MODE to set the
respective sinlge sweep. You can alternatively use an IMMediate command instead of
the respective SWEep:[FREQ:]|POW:EXECute command.
Example:
TRIG
starts all active sweeps.
Usage: Event
Manual operation: See "Execute Single Sweep - LF Sweep" on page 191
7.17 UNIT Subsystem
The UNIT subsystem contains the commands specifying which units are valid if no unit
is indicated in a command. These settings are valid for the entire instrument.
:UNIT:ANGLe <Angle>
Defines the default unit for the phase modulation angle. It is not valid for commands which
determine angle values, e.g. RF phase. It does not influence the manual control param-
eter unit and the display.
Parameters:
<Angle> DEGRee | RADian
*RST: RADian
UNIT Subsystem

Table of Contents