Remote Control Commands
R&S
®
SMB100A
246Operating Manual 1407.0806.32 ─ 13
Sets parallel poll enable register to the indicated value. The query returns the contents
of the parallel poll enable register in decimal form.
Parameters:
<Value> Range: 0 to 255
*PSC <Action>
Power on Status Clear
Determines whether the contents of the ENABle registers are preserved or reset when
the instrument is switched on. Thus a service request can be triggered when the instru-
ment is switched on, if the status registers ESE and SRE are suitably configured. The
query reads out the contents of the "power-on-status-clear" flag.
Parameters:
<Action> 0 | 1
0
The contents of the status registers are preserved.
1
Resets the status registers.
*RCL <Number>
ReCaLl calls up the instrument settings from an intermediate memory identified by the
specified number. The instrument settings can be stored to this memory using the com-
mand *SAV with the associated number.
It also activates the instrument settings which are stored in a file and loaded using the
MMEMory:LOAD <number>, <file_name.extension> command.
Manual operation:
See "Recall" on page 110
*RST
ReSeT
Sets the instrument to a defined default status. The default settings are indicated in the
description of commands.
The command is equivalent to SYSTem:PRESet.
Usage:
Setting only
Manual operation: See "Preset" on page 80
*SAV <Number>
SAVe stores the current instrument settings under the specified number in an intermedi-
ate memory. The settings can be recalled using the command *RCL with the associated
number.
Common Commands