7.12.12 SOURce:POWer Subsystem
This subsystem contains the commands for setting the output level, level control and level
correction of the RF signal.
Other units can also be used instead of dBm:
by entering the unit directly after the numerical value (example :POW 0.5V)
by changing the DEFault unit in the UNIT system (see the command :UNIT:
POWer).
[:SOURce<hw>]:POWer:ALC:OMODe.............................................................................339
[:SOURce<hw>]:POWer:ALC:SONCe..............................................................................340
[:SOURce<hw>]:POWer:ALC[:STATe].............................................................................340
[:SOURce<hw>]:POWer:ATTenuation:RFOFf:MODE.........................................................340
[:SOURce<hw>]:POWer[:LEVel][:IMMediate][:AMPLitude]..................................................341
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:OFFSet........................................................342
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:RCL.............................................................342
[:SOURce<hw>]:POWer:LIMit[:AMPLitude].......................................................................343
[:SOURce<hw>]:POWer:LMODe.....................................................................................343
[:SOURce<hw>]:POWer:MANual.....................................................................................344
[:SOURce<hw>]:POWer:MODE.......................................................................................345
[:SOURce<hw>]:POWer:POWer......................................................................................345
[:SOURce<hw>]:POWer:SPC:CRANge............................................................................346
[:SOURce<hw>]:POWer:SPC:DELay...............................................................................346
[:SOURce<hw>]:POWer:SPC:PEAK................................................................................346
[:SOURce<hw>]:POWer:SPC:SELect..............................................................................347
[:SOURce<hw>]:POWer:SPC:STATe...............................................................................347
[:SOURce<hw>]:POWer:SPC:TARGet.............................................................................347
[:SOURce<hw>]:POWer:STARt.......................................................................................347
[:SOURce<hw>]:POWer:STEP[:INCRement]....................................................................348
[:SOURce<hw>]:POWer:STEP:MODE.............................................................................348
[:SOURce<hw>]:POWer:STOP........................................................................................349
[:SOURce<hw>]:POWer:ALC:OMODe <OffMode>
The command sets the level control mode which becomes active when automatic level
control is deactivated (ALC Off).
Parameters:
<OffMode> SHOLd
SHOLd
Level control is activated briefly if the level or frequency changes
("ALC Off Sample & Hold").
*RST: SHOLd
Example:
POW:ALC OFF
deactivates automatic level control for RF output A.
POW:ALC:OMOD SHOL
level control is briefly activated if the frequency or level changes.