Remote Control Commands
R&S
®
SMB100A
340Operating Manual 1407.0806.32 ─ 13
[:SOURce<hw>]:POWer:ALC:SONCe
The command briefly activates level control for correction purposes.
Example:
POW:ALC OFF
deactivates automatic level control for RF output A.
POW:ALC:SONC
level control is performed once only.
Usage: Event
Manual operation: See "Search Once - ALC" on page 130
[:
SOURce<hw>]:POWer:ALC[:STATe] <State>
The command activates/deactivates automatic level control.
Parameters:
<State> ON | OFF | AUTO
ON
Internal level control is permanently activated.
OFF
Internal level control is deactivated; Sample & Hold mode is acti-
vated.
AUTO
Internal level control is activated/deactivated automatically
depending on the operating state.
*RST: AUTO
Example:
POW:ALC ON
activates automatic level control for RF output A.
Manual operation: See "State - ALC" on page 129
[:SOURce<hw>]:POWer:ATTenuation:RFOFf:MODE <Mode>
Selects the attenuator mode, when the RF signal is switched off.
The setting of the RF OFF mode is not affected by an instrument preset (PRESET key),
*RST and the "Save/Recall" function. This parameter is influenced only by the Factory
Preset.