Remote Control Commands
R&S
®
SMB100A
328Operating Manual 1407.0806.32 ─ 13
Parameters:
<Frequency> <Frequency#1>{, <Frequency#2>, ...} | block data
The data can be given either as a list of numbers (list can be of
any length and list entries must be separated by commas) or as
binary block data. When block data is transferred, 8 (4) bytes are
always interpreted as a floating-point number with double accu-
racy (see :FORMat[:DATA] on page 257).
Range: 300 kHz to RFmax (RFmax depending on model)
Example:
LIST:SEL '/var/user//Lists/Listmode/list3'
selects list3 for editing. List3 is created if it does not yet exist.
SOUR:LIST:FREQ 1.4GHz, 1.3GHz, 1.2GHz,...
specifies the frequency values in list3. If the list already contains
data, it is overwritten.
Usage: SCPI confirmed
Manual operation: See "Edit List Mode Data... - List Mode" on page 163
[:
SOURce<hw>]:LIST:FREQuency:POINts?
The command queries the length (in points) of the FREQuency component of the selected
list.
Return values:
<Points> integer
Range: 0 to INT_MAX
Increment: 1
*RST: 0
Example:
LIST:SEL '/var/user//Lists/Listmode/list3'
selects list3 for editing. List3 is created if it does not yet exist.
LIST:FREQ:POIN?
queries the number of frequency values in list3.
Response: 327
list3 has 327 frequency entries.
Usage: Query only
[:SOURce<hw>]:LIST:INDex <Index>
The command sets the list index in step mode (LIST:MODE STEP).
After the trigger signal the frequency and level settings of the selected index are pro-
cessed in List mode.
Parameters:
<Index> integer
Increment: 1
*RST: 0
SOURce Subsystem