Remote Control Commands
R&S
®
SMB100A
248Operating Manual 1407.0806.32 ─ 13
Usage: Event
7.2 Preset Commands
The preset commands are not bundled in one subsystem. Therefore, they are listed sep-
arately in this section. In addition, a specific preset command is provided for each digital
standard and for the fader. These specific commands are described in the associated
subsystems.
Four presetting actions are available:
● Activating the default state of all internal instrument functions (*RST on page 246).
Functions that concern the integration of the instrument into a measurement setup
are not changed, e.g. GPIB address or reference oscillator source settings.
● Activating the preset state of the parameters related to the selected signal path (:
SOURce<hw>:PRESet on page 248)
● Activating the preset state of all parameters that are not related to the signal path (
:DEVice:PRESet on page 248)
● Activating the original state of delivery (factory reset, :SYSTem:FPReset
on page 249). Only functions that are protected by a password remain unchanged
as well as the passwords themselves.
:DEVice:PRESet
The command presets all parameters which are not related to the signal path, including
the LF generator.
Example:
DEV:PRES
presets all instruments settings that are not related to the signal
path
Usage: Event
:SOURce<hw>:PRESet
The command presets all parameters which are related to the selected signal path.
Fading (if available) and transient recorder are only preset by command *RST.
Example:
SOUR:PRES
presets all settings that are related to signal path
:SYSTem:PRESet
The command triggers an instrument reset. It has the same effect as:
●
the PRESET key on the front panel
●
the *RST command
Preset Commands