:DISPlay:PSAVe:HOLDoff <HoldoffTimeMin>
Sets the waiting time for the screen-save mode of the display. The entry is dimensionless.
Parameters:
<HoldoffTimeMin> integer
Range: 1 to 60
*RST: 10
Default unit: minute
Example:
DISP:PSAV:HOLD 8
sets the timeout of the screen saver to 8 minutes.
Manual operation: See "Wait Time" on page 93
:
DISPlay:PSAVe[:STATe] <State>
Activates the screen-save mode of the display. If activated, the display including backlight
is completely switched off after the elapse of the wait time, provided no entries via front
panel, external mouse or external keyboard are made. To determine the wait time, use
command :DISPlay:PSAVe:HOLDoff.
This mode is recommended for protecting the display, especially if the instrument is
operated via remote control.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: OFF
Example:
DISP:PSAV ON
activates screen saver mode.
Manual operation: See "Screen Saver Active" on page 92
:DISPlay:UPDate <Update>
Activates the refresh mode of the display.
Parameters:
<Update> 0 | 1 | OFF | ON
*RST: 1
Example:
DISP:UPD ON
activates automatic update of the display at defined time intervals.
7.6 FORMat Subsystem
The FORMat subsystem contains the commands which determine the format of the data
that the R&S SMB returns to the controller. This affects all query commands which return