Remote Control Commands
R&S
®
SMB100A
247Operating Manual 1407.0806.32 ─ 13
To transfer the stored instrument settings in a file, use the command :MMEMory:
STORe:STATe.
Manual operation: See "Save" on page 109
*SRE <Contents>
Service Request Enable
Sets the service request enable register to the indicated value. This command determines
under which conditions a service request is triggered.
Parameters:
<Contents> Contents of the service request enable register in decimal form.
Bit 6 (MSS mask bit) is always 0.
Range: 0 to 255
*STB?
STatus Byte query
Reads the contents of the status byte in decimal form.
Usage:
Query only
*TRG
TRiGger
Triggers all actions waiting for a trigger event. In particular, *TRG generates a manual
trigger signal (Manual Trigger). This common command complements the commands of
the TRIGger subsystem.
Usage:
Event
*TST?
self TeST query
Triggers selftests of the instrument and returns an error code in decimal form (see Service
Manual supplied with the instrument). "0" indicates no errors occured.
Usage:
Query only
*WAI
WAIt to continue
Prevents servicing of the subsequent commands until all preceding commands have
been executed and all signals have settled (see also command synchronization and
*OPC).