7.12.9 SOURce:PGEN Subsystem
This subsystem contains the commands for setting the pulse generator.
[:SOURce<hw>]:PGENerator:STATe...............................................................................334
[:SOURce<hw>]:PGENerator:STATe <State>
The command activates/deactivates the output of the video/sync signal at the PULSE
VIDEO connector at the rear of the instrument.
The signal output and the pulse generator are automatically switched on with activation
of pulse modulation if pulse generator is selected as modulation source. The signal output
can be switched off subsequently.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: OFF
Example:
PULM:SOUR INT
selects the internal pulse generator as source for pulse modulation
PULM:STAT ON
activates pulse modulation. The pulse generator and the output of
the signals at the PULSE VIDEO connector are automatically acti-
vated in addition.
PGEN:STAT OFF
deactivates the output of the pulse signal by the pulse generator
at the PULSE VIDEO connector. The pulse modulation of the RF
carrier must be activated with command SOURce:PULM:STATe.
Manual operation: See "Video Sync Signal State - Pulse Generator" on page 195
7.12.10 SOURce:PHASe Subsystem
This subsystem contains the commands for adjusting the phase of the RF output signal
relative to a reference signal of the same frequency.
[:SOURce<hw>]:PHASe.................................................................................................334
[:SOURce<hw>]:PHASe:REFerence................................................................................335
[:SOURce<hw>]:PHASe <Phase>
The command specifies the phase variation relative to the current phase. The variation
can be specified in RADians.