Remote Control Commands
R&S
®
SMB100A
361Operating Manual 1407.0806.32 ─ 13
Usage: Query only
Options: R&S SMB-K27 (Pulse Train)
Manual operation: See "Select ASCII Source / Destination - Import/Export Pulse
Train Files" on page 201
[:SOURce<hw>]:PULM:TRAin:DEXChange:AFILe:EXTension <Extension>
The command selects the file extension of the ASCII file to be imported or exported.
Selection TXT (text file) or CSV (Excel file) is available.
Parameters:
<Extension> TXT | CSV
*RST: TXT
Example:
MMEM:CDIR 'var/lists/pulsetrain/import'
selects the directory for the ASCII files with ontime/offtime/repeti-
tion values.
PULM:TRA:DEXC:AFIL:EXT TXT
selects that ASCII files with extension *.txt are listed.
PULM:TRA:DEXC:AFIL:CAT?
queries the available files with extension *.txt.
Response: 'train1','train2
the ASCII files train1.txt and train2.txt are available.
Options: R&S SMB-K27 (Pulse Train)
Manual operation: See "Extension – ASCII File Settings" on page 200
[:SOURce<hw>]:PULM:TRAin:DEXChange:AFILe:SELect <Filename>
The command selects the ASCII file to be imported or exported.
The ASCII files are stored with the fixed file extensions *.txt or *.csv in a directory of
the user's choice. The directory applicable to the commands is defined with the command
MMEMory:CDIR.
Parameters:
<Filename> string
Example:
MMEM:CDIR 'var/lists/pulsetrain/import'
selects the directory for the ASCII files with ontime/offtime/repeti-
tion values.
PULM:TRA:DEXC:MODE IMP
selects that ASCII files with ontime/offtime/repetition values are
imported and transferred into pulse train lists.
PULM:TRA:DEXC:AFIL:SEL 'train.csv'
selects that ASCII file train.csv is imported.
PULM:TRA:DEXC:SEL 'train_imp'
selects that the ASCII file train.csv is imported into pulse train
list train_imp.