Instrument Function
R&S
®
SMB100A
109Operating Manual 1407.0806.32 ─ 13
Directory
Selects the directory in which the file is to be stored.
The window opposite lists all settings files in this directory.
A new directory can be created in the "File Manager" dialog.
Remote command:
:MMEMory:CDIRectory on page 270
File List
Displays the files which are in the selected directory.
If a file is highlighted, it is overwritten when the file is stored.
Remote command:
:MMEMory:CATalog? on page 269
File Name
Enter the file name of the file without file extension. This file is then created.
Remote command:
n.a.
Save
Stores the current instrument settings under the specified path.
Remote command:
*SAV on page 246
:MMEMory:STORe:STATe on page 274
Save Immediate x
Stores the current instrument setting in one of the three intermediate memories.
These instrument settings are retained until a different instrument setting is stored in the
intermediate memory. When the instrument is switched off, the contents of the inter-
mediate memories are retained.
Remote command:
*SAV on page 246
File Manager
Calls the "File Management" menu.
Directories can be created and files managed in this menu (see chapter 5.2.8.4, "File
Manager", on page 111).
Remote command:
n.a.
5.2.8.3 Loading Instrument Settings
If "Recall" is selected under "Select Operation", the "File" menu provides options for
loading complete instrument settings. Here it is possible to select whether the current or
stored frequency and level settings are to be used.
General Instrument Settings