Remote Control Commands
R&S
®
SMB100A
285Operating Manual 1407.0806.32 ─ 13
Example:
SENS:SOUR USER
selects user-defined source.
SENS:FREQ 2.44 GHz
enters the RF frequency of the source which is 2.44 GHz.
Manual operation: See "Frequency - Power Sensors" on page 145
:SENSe<ch>[:POWer]:OFFSet <Offset>
The command enters a level offset which is added to the measured level value after
activation with command SENSe[:POWer]:OFFSet:STATe ON. This allows e.g. an
attenuator in the signal path to be considered.
Parameters:
<Offset> float
Range: -100.0 to 100.0
*RST: 0
Default unit: dB
Example:
SENS:POW:OFFS 10.0
sets a level offset of 10 dB
Manual operation: See "Level Offset - Power Sensors" on page 145
:SENSe<ch>[:POWer]:OFFSet:STATe <State>
The command activates the addition of the level offset to the measured value. The level
offset value is set with command SENSe[:POWer]:OFFSet.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: OFF
Example:
SENS1:POW:OFFS 0.4dB
sets a level offset of 0.4 dB
SENS1:POW:OFFS:STAT ON
a level offset of 0.4 dB is added to the measured value.
Manual operation: See "Level Offset - Power Sensors" on page 145
SENSe<ch>[:POWer]:SNUMber?
The command queries the serial number of the sensor.
Return values:
<Snumber> string
Example:
SENS:SNUM?
queries the serial number.
Usage: Query only
Manual operation: See "Current Sensors" on page 87