EasyManua.ls Logo

Rohde & Schwarz SMB100A - [: S Ource<Hw>]: P ULM: S Tate; [: S Ource<Hw>]: P ULM: T Rain: D Elete; [:Source<Hw>]:Pulm:train:catalog

Rohde & Schwarz SMB100A
481 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMB100A
354Operating Manual 1407.0806.32 ─ 13
Example:
PULM:SOUR INT
selects the internal modulation source.
PULM:STAT ON
activates the pulse modulation.
Usage: SCPI confirmed
Options: R&S SMB-K21 or R&S SMB-K22 (Pulse Modulator)
Manual operation: See "Source" on page 180
[:SOURce<hw>]:PULM:STATe <State>
Activates/deactivates the pulse modulation.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: OFF
Example:
PULM:STAT ON
activates pulse modulation.
Options: R&S SMB-K21 or R&S SMB-K22 (Pulse Modulator)
Manual operation: See "State" on page 180
[:SOURce<hw>]:PULM:TRAin:CATalog?
Queries a list of available pulse train files. The individual pulse train files are separated
by commas.
The files are stored with the fixed file extensions *.pulstrn in a directory of the user's
choice. The directory applicable to the commands is defined with the command
MMEMory:CDIR.
Return values:
<Catalog> string
Example:
MMEM:CDIR '/var/user/lists/pulsetrain'
selects the directory for the pulse train files.
PULM:TRA:CAT?
queries the available files.
Response: 'P_CONS','P_INCR','P_DECR'
the lists P_CONS, P_INCR and P_DECR are available.
Usage: Query only
Options: R&S SMB-K27 (Pulse Train)
Manual operation: See "Pulse Train Data – Pulse Generator" on page 198
[:SOURce<hw>]:PULM:TRAin:DELete <Filename>
Deletes the specified pulse train file.