Instrument Function
R&S
®
SMB100A
198Operating Manual 1407.0806.32 ─ 13
A pulse train is a sequence of pulses with user-defined on and off times. The ontime/
offtime value pairs are defined in a pulse train list. The currently used pulse train file is
displayed in the sub menu.
An internally created pulse train list can be exported or externally created list can be
imported.
Pulse Train Data – Pulse Generator
Calls the "File Select" menu for selecting and creating a pulse train file or to open the
"File Manager".
Remote command:
[:SOURce<hw>]:PULM:TRAin:CATalog? on page 354
[:SOURce<hw>]:PULM:TRAin:SELect on page 358
[:SOURce<hw>]:PULM:TRAin:DELete on page 354
Edit Pulse Train Data – Pulse Generator
Calls the Pulse Train Dialog.
Pulse Train Dialog
The pulse train defined by the selected file is displayed.
The display of the pulse train can be zoomed, the zoom center is defined by the blue
marker.
"Edit"
Calls the Edit Pulse Train Data dialog for editing the pulse train.
"Zoom Position"
Positions the blue marker in the pulse train. The marker defines the
center of any zoom in or zoom out action.
"Zoom In/Out"
Zooms in or out the graphical display of the pulse train. The zoom factor
of each zoom action is 2.
Edit Pulse Train Data
(For Pulse Mode Train only)
Modulation Generator and LF Output