Instrument Function
R&S
®
SMB100A
199Operating Manual 1407.0806.32 ─ 13
Calls the editor for editing the ontime/offtime values of the selected file. A file consists of
any number of ontime/offtime value pairs. The currently selected file is displayed.
"Ontime/us"
Sets the ontime of the ontime/offtime value pair.
Remote command:
[:SOURce<hw>]:PULM:TRAin:ONTime on page 356
"Offtime/us"
Sets the offtime of the ontime/offtime value pair.
Remote command:
[:SOURce<hw>]:PULM:TRAin:OFFTime on page 355
"Count"
Sets the number of repetitions for each ontime/offtime value pair.
Tip: If you set "Count = 0", the corresponding value pair is ignored in
the pulse sequence. With this function you can omit value pairs indi-
vidually, without deleting them from the table. To activate them again
set a value other than 0.
Remote command:
[:SOURce<hw>]:PULM:TRAin:REPetition on page 357
"Goto"
Selects row for editing.
If Goto row is selected, a window opens for entering the requested row.
(it is not possible to change individual positions of the list)
"Edit"
Calls a selection of possible actions described below.
"Insert Row"
Insert a new row before the marked row.
"Insert Range"
Insert new rows before the marked row. The number of rows to be
inserted can be defined in an entry window.