Remote Control Commands
R&S
®
SMB100A
279Operating Manual 1407.0806.32 ─ 13
Up to four sensors can be connected to the signal generator. They are distinguished by
means of the suffix under SENSe, i.e. SENSe [1] ... SENSe 4.
:INITiate<ch>[:POWer]:CONTinuous................................................................................279
:READ<ch>[:POWer]?....................................................................................................279
:SENSe<ch>[:POWer]:CORRection:SPDevice:STATe.......................................................280
:SENSe<ch>[:POWer]:DISPlay:PERManent:STATe..........................................................280
:SENSe<ch>[:POWer]:DISPlay:PERManent:PRIority.........................................................281
:SENSe<ch>[:POWer]:FILTer:LENGth:AUTO?..................................................................281
:SENSe<ch>[:POWer]:FILTer:LENGth[:USER]..................................................................281
:SENSe<ch>[:POWer]:FILTer:NSRatio.............................................................................282
:SENSe<ch>[:POWer]:FILTer:NSRatio:MTIMe..................................................................282
:SENSe<ch>[:POWer]:FILTer:SONCe..............................................................................283
:SENSe<ch>[:POWer]:FILTer:TYPE................................................................................283
:SENSe<ch>[:POWer]:LOGGing:STATe...........................................................................284
:SENSe<ch>[:POWer]:FREQuency..................................................................................284
:SENSe<ch>[:POWer]:OFFSet........................................................................................285
:SENSe<ch>[:POWer]:OFFSet:STATe.............................................................................285
SENSe<ch>[:POWer]:SNUMber?....................................................................................285
:SENSe<ch>[:POWer]:SOURce.......................................................................................286
SENSe<ch>[:POWer]:STATus[:DEVice]?.........................................................................286
SENSe<ch>[:POWer]:SVERsion?....................................................................................286
SENSe<ch>[:POWer]:TYPE?..........................................................................................286
:SENSe<ch>[:POWer]:ZERO..........................................................................................287
:SENSe<ch>:UNIT[:POWer]............................................................................................287
:INITiate<ch>[:POWer]:CONTinuous <Continuous>
The command switches the local state of the continuous power measurement by the R&S
NRP-Zxx power sensors on and off. Switching the local state off enhances the measure-
ment performance during remote control
The remote measurement is triggered by the READ query (command :READ<ch>[:
POWer]? on page 279) which also provides the measurement results. The local state is
not influenced by this command, measurements results can be retrieved with local state
on or off.
Parameters:
<Continuous> 0 | 1 | OFF | ON
*RST: OFF
Example:
INIT:CONT ON
switches local state of continuous power measurement on.
Manual operation: See "State - Power Sensors" on page 143
:READ<ch>[:POWer]?
The command triggers the measurement with power sensors and provides the power
measurement result of the selected power sensor. The value is provided with the unit set
with command SENSe:UNIT[:POWer].
SENSe, READ and INITiate Subsystems