Example: Commands and queries in one message
The response to a query combined in a program message with commands that affect the
queried value is not predictable.
The following commands always return the specified result:
:FREQ:STAR 1GHZ;SPAN 100 :FREQ:STAR?
Result:
1000000000 (1 GHz)
Whereas the result for the following commands is not specified by SCPI:
:FREQ:STAR 1GHz;STAR?;SPAN 1000000
The result could be the value of STARt before the command was sent since the instru-
ment might defer executing the individual commands until a program message terminator
is received. The result could also be 1 GHz if the instrument executes commands as they
are received.
As a general rule, send commands and queries in different program messages.
Example: Overlapping command with *OPC
The instrument implements INITiate[:IMMediate] as an overlapped command.
Assuming that INITiate[:IMMediate] takes longer to execute than *OPC, sending
the following command sequence results in initiating a sweep and, after some time, set-
ting the OPC bit in the ESR:
INIT; *OPC.
Sending the following commands still initiates a sweep:
INIT; *OPC; *CLS
However, since the operation is still pending when the instrument executes *CLS, forcing
it into the "Operation Complete Command Idle" State (OCIS), *OPC is effectively skipped.
The OPC bit is not set until the instrument executes another *OPC command.
6.4.1 Preventing Overlapping Execution
To prevent an overlapping execution of commands, one of the commands *OPC, *OPC?
or *WAI can be used. All three commands cause a certain action only to be carried out
after the hardware has been set. By suitable programming, the controller can be forced
to wait for the corresponding action to occur.