Remote Control Commands
R&S
®
SMB100A
278Operating Manual 1407.0806.32 ─ 13
Example:
OUTP:PROT:TRIP
Queries the state of the protective circuit for RF output A.
Response: 0
The protective circuit has not tripped.
Response: 1
The protective circuit has tripped.
Usage: Query only
Manual operation: See "Overload" on page 127
:OUTPut<hw>[:STATe] <State>
This command activates and deactivates the RF output.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
OUTP OFF
deactivates the RF output.
Manual operation: See "State RF" on page 115
:OUTPut<hw>[:STATe]:PON <Pon>
This command selects the state which the RF output assumes when the instrument is
switched on.
Parameters:
<Pon> OFF | UNCHanged
OFF
The output is deactivated when the instrument is switched on.
UNCHanged
When the instrument is switched on, the output remains in the
same state as it was when the instrument was switched off.
*RST: UNCHanged
Example:
OUTP:PON OFF
RF output A is deactivated when the instrument is switched on.
Manual operation: See "Power-On State - RF Signal" on page 128
7.11 SENSe, READ and INITiate Subsystems
The SENSe subsystem contains the commands for configuring the power measurements
with R&S NRP-Zxx power sensor(s) connected to the generator. The measurement is
started and the measurement result retrieved with the READ command. The description
of this commands is included in the following.
SENSe, READ and INITiate Subsystems