Remote Control Commands
R&S
®
SMB100A
277Operating Manual 1407.0806.32 ─ 13
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
OUTP:FILT:AUTO 0
OUTP:FILT:LPAS:STAT 1
Selects manual mode and activates the low harmonic filter.
OUTP:FILT:STAT?
Queries the current filter activity.
Manual operation: See "State" on page 125
:
OUTPut<hw>:IMPedance?
The command queries the impedance of the RF outputs. This permits converting the
output level between units V and W. The impedances cannot be changed.
Return values:
<Impedance> G1K | G50 | G10K
*RST: G50
Default unit: Ohm
Example:
OUTP:IMP
queries the impedance of RF output.
Response: 50
the impedance is 50 ohms
Usage: Query only
:OUTPut<hw>:PROTection:CLEar
Resets the protective circuit after it has been tripped. The state of the output is again
determined by OUTPut:STATe.
Example:
OUTP:PROT:CLE
resets the protective circuit for RF output.
Usage: Event
Manual operation: See "Overload" on page 127
:OUTPut<hw>:PROTection:TRIPped?
The command queries the state of the protective circuit.
Return values:
<Tripped> 0 | 1 | OFF | ON
*RST: 0
OUTPut Subsystem