Remote Control Commands
R&S
®
SMB100A
276Operating Manual 1407.0806.32 ─ 13
:OUTPut<hw>:ALC:SEARch:MODE <Mode>
The command activates/deactivates the RF output during the power search.
Parameters:
<Mode> NORMal | MINimum
*RST: OFF
Example:
POW:ALC:SEAR:MODE NORM
during the power search, the RF output is active.
Manual operation: See "RF During Power Search - ALC" on page 130
:OUTPut<hw>:AMODe <AMode>
The command switches the mode of the attenuator at the RF output (Attenuator MODe).
Parameters:
<AMode> AUTO | FIXed
AUTO
The attenuator is switched automatically. The level settings are
made in the full range.
FIXed
The level settings are made without switching the attenuator.
When this operating mode is switched on, the attenuator is fixed
in its current position and the resulting variation range is defined.
*RST: AUTO
Example:
POW:ALC ON
activates automatic level control for RF output.
OUTP:AMOD FIX
sets the fixed mode with uninterrupted level for RF output.
Manual operation: See "Attenuator Mode" on page 125
:OUTPut<hw>:FILTer:AUTO <Auto>
Activates automatic switching of the low harmonic filter.
Parameters:
<Auto> 0 | 1 | ON | OFF
*RST: 0
Example:
OUTP:FILT:AUTO 1
Activates the auto mode.
Manual operation: See "Mode" on page 125
:OUTPut<hw>:FILTer[:LPASs]:STATe <State>
Switches the filter state in manual mode and disables the automatic mode, if activated.
OUTPut Subsystem