EasyManua.ls Logo

Rohde & Schwarz SMB100A - [: S Ource<Hw>]: S Weep[: F Requency]: L Fconnector; [: S Ource<Hw>]: S Weep[: F Requency]: M ODE

Rohde & Schwarz SMB100A
481 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMB100A
379Operating Manual 1407.0806.32 ─ 13
Example:
TRIG:FSW:SOUR SING
SOUR:SWE:FREQ:MODE AUT
SWE:FREQ:EXEC
triggers a frequency sweep at the RF output.
Usage: Event
Manual operation: See "Execute Single Sweep - Frequency Sweep" on page 151
[:SOURce<hw>]:SWEep[:FREQuency]:LFConnector <LfConnector>
The command activates the output of a sweep voltage ramp at the LF connector.
The voltage range is set with commands
SOURce:SWEep:FREQuency:OVOLtage:STARt and …:STOP
Parameters:
<LfConnector> OFF | ON
*RST: OFF
Example:
SWE:LFC ON
activates the output of a linear voltage ramp from sweep start to
sweep stop at the LF connector.
SWE:OVOL:STAR 0V
SWE:OVOL:STOP 3V
'the voltage at sweep start is 0 Volt and at sweep stop 3 V.
Manual operation: See "Use LF connector to output sweep voltage - RF Frequency
Sweep" on page 154
[:SOURce<hw>]:SWEep[:FREQuency]:MODE <Mode>
Sets the sweep mode.
The keyword [:FREQuency] can be omitted (see example). The command is then
SCPI-compliant.
SOURce Subsystem

Table of Contents

Other manuals for Rohde & Schwarz SMB100A

Related product manuals