Instrument Function
R&S
®
SMB100A
93Operating Manual 1407.0806.32 ─ 13
If activated, the display including backlight is completely switched off after the elapse of
the "Wait Time" when no entries via front panel, external mouse or external keyboard are
made.
This mode is recommended for preserving the display especially if the instrument is
exclusively operated via remote control.
Remote command:
ïŠ :ïŠ DISPlay:ïŠ PSAVe[:ïŠ STATe]ïŠ on page 256
Wait Time
Enters the idle time that must elapse before the display lamp is shut off when no entries
are made.
Remote command:
ïŠ :ïŠ DISPlay:ïŠ PSAVe:ïŠ HOLDoffïŠ on page 256
Toggle Summary Screen
Activates/deactivates the magnified frequency and level indication. If activated, the fre-
quency and level indication covers the complete display.
Remote command:
n.a.
Layout (Keyboard)
Selects the keyboard layout for the selected keyboard language.
The assignment of some keys depends on the selected layout and language.
Remote command:
ïŠ :ïŠ KBOard:ïŠ LAYoutïŠ on page 266
ïŠ :ïŠ KBOard:ïŠ LANGuageïŠ on page 265
5.2.3.13 Remote Channel Settings
The "Remote Channel Settings" dialog provides access to the settings for remote control.
The dialog is opened using the SETUP or MENU key under "Remote".
Besides the standard GPIB-setting, the selection of one of the possible emulations and
a user defined definition of the *IDN and *OPT string is possible.
General Instrument Settings