[:SOURce<hw>]:SWEep[:FREQuency]:DWELl..................................................................378
[:SOURce<hw>]:SWEep[:FREQuency]:EXECute..............................................................378
[:SOURce<hw>]:SWEep[:FREQuency]:LFConnector.........................................................379
[:SOURce<hw>]:SWEep[:FREQuency]:MODE..................................................................379
[:SOURce<hw>]:SWEep[:FREQuency]:OVOLtage:STARt..................................................380
[:SOURce<hw>]:SWEep[:FREQuency]:OVOLtage:STOP...................................................381
[:SOURce<hw>]:SWEep[:FREQuency]:POINts.................................................................381
[:SOURce<hw>]:SWEep[:FREQuency]:RUNNing?............................................................382
[:SOURce<hw>]:SWEep[:FREQuency]:SHAPe.................................................................382
[:SOURce<hw>]:SWEep[:FREQuency]:SPACing...............................................................382
[:SOURce<hw>]:SWEep[:FREQuency]:STEP[:LINear].......................................................383
[:SOURce<hw>]:SWEep[:FREQuency]:STEP:LOGarithmic................................................384
[:SOURce<hw>]:SWEep:POWer:DWELl..........................................................................384
[:SOURce<hw>]:SWEep:POWer:EXECute.......................................................................385
[:SOURce<hw>]:SWEep:POWer:MODE...........................................................................385
[:SOURce<hw>]:SWEep:POWer:POINts..........................................................................386
[:SOURce<hw>]:SWEep:POWer:RUNNing?.....................................................................386
[:SOURce<hw>]:SWEep:POWer:SHAPe..........................................................................386
[:SOURce<hw>]:SWEep:POWer:SPACing:MODE?...........................................................387
[:SOURce<hw>]:SWEep:POWer:STEP[:LOGarithmic].......................................................387
[:SOURce<hw>]:SWEep:RESet[:ALL]..............................................................................387
[:SOURce<hw>]:SWEep[:FREQuency]:DWELl <Dwell>
The command sets the time taken for each frequency step of the sweep.
The keyword [:FREQuency] can be omitted (see example). The command is then SCPI-
compliant.
Tip: It is recommended to switch off the "GUI Update" for optimum sweep performance
especially with short dwell times (SYSTem:DISPlay:UPDate OFF).
Parameters:
<Dwell> float
Range: 5 ms to 100 s
Increment: 100 us
*RST: 10 ms
Default unit: s
Example:
SWE:DWEL 12 ms
sets a dwell time of 12 ms for a frequency sweep at the RF output.
Manual operation: See "Dwell Time - Frequency Sweep" on page 154
[:SOURce<hw>]:SWEep[:FREQuency]:EXECute
Starts an RF frequency sweep cycle manually.
The command is only effective in single mode.