Remote Control Commands
R&S
®
SMB100A
282Operating Manual 1407.0806.32 ─ 13
Parameters:
<User> float
Range: 1 to 65536
*RST: 1
Example:
SENS:FILT:TYPE USER
selects user filter mode.
SENS:FILT:LENG 16
sets a filter length of 16. The resulting measurement time is 640
ms (2x16x20 ms).
Manual operation: See "Filter Length - Power Sensors" on page 145
:
SENSe<ch>[:POWer]:FILTer:NSRatio <NSRatio>
The command defines the noise content for fixed noise filter mode
(:SENSe<[1]...3>:POWer:FILTer:TYPE NSRatio). This value determines the
proportion of intrinsic noise in the measured result.
Parameters:
<NSRatio> float
Range: 0.001 to 1
Increment: 0.001
*RST: 0.01
Example:
SENS1:FILT:TYPE NSR
selects fixed noise filter mode for the power sensor connected to
the SENSOR connector.
SENS1:FILT:NSR 0.2
sets a noise content of 0.2.
Manual operation: See "Filter Length - Power Sensors" on page 145
:SENSe<ch>[:POWer]:FILTer:NSRatio:MTIMe <MTime>
The command defines the timeout for fixed noise filter mode
(:SENSe<[1]...3>:POWer:FILTer:TYPE NSRatio). This value ensures limited
settling times.
Parameters:
<MTime> float
Range: 1 to 999.99
Increment: 0.01
*RST: 4
Default unit: s
SENSe, READ and INITiate Subsystems